The Gazette, January 2024: active or settle plates in grade A
EU GMP Annex 1 (clause 9.24) asks for continuous viable air monitoring in grade A for the full duration of critical processing, including equipment assembly. This issue compares active air sampling and settle plates in two tests.
The tests
In the first test an AIRBIO TRIO sampled 100 l/min in three one-hour sessions (6,000 litres per session), while three settle plates stayed open for one hour each.
Active samplingSettle plate
In the second test a single settle plate on the TRIO.SETTLE floor stand stayed open for three hours next to the three plates of the active sampler. The document reports 162 CFU for active sampling and 23 for the settle plate.
Two different measurements
Active sampling counts the microorganisms in a known volume of air, usually 1 m³. A settle plate counts what lands on its surface. In a closed space such as an isolator, people do not affect the result; with operators present, their movements make deposition on the agar uneven.
Across the two tests, settle plates counted 86% to 89% fewer CFU. The Gazette says the choice of method belongs in the contamination control strategy (CCS), with its reasons.
Results at a glance
- First test: 162 CFU with active sampling, 18 with settle plates (89% fewer).
- Second test: 162 CFU against 23 on a single plate left open for three hours (86% fewer).
The Gazette also notes that the settle plate method cannot be validated. For a real evaluation it recommends active sampling, and asks for these data to be taken into account in the contamination control strategy.
Grade A in 2024
- Gazette, JanActive or settle plates
- Gazette, MarTwo heads
- Release, NovShifts of 3, 6 or 9 hours
- Gazette, Nov2 and 3 heads
- Release, DecFilling lines and isolators
- Release, DecModels for the cleanroom
The Gazette in 2024
- JanActive or settle plates in grade A
- FebWhere to sample in a cleanroom
- MarWhy two heads
- AprAIRBIO ONE RAPID-VIRUS
- MayThe sterile Daily Shift head
- JunThe risk of bioaerosols
- JulA short course in microbiology
- AugCleanroom tips
- SepHow to enter a cleanroom
- OctDetecting viruses fast
- Nov2- and 3-head samplers
- DecActive or passive?
TRIO.BAS in 2024
- JanuaryThe Gazette restarts with grade A
- FebruarySales staff at Villa Cella
- AprilExpoFarma, Mexico
- AprilKorea Lab, South Korea
- MayCleanroom Technology Conference, UK
- JuneAFI Symposium, Rimini
- OctoberRelease: active or passive?
- OctoberIran Pharma, Tehran
- OctoberPharma Microbiology, Sofia
- OctoberWorkshop in Anápolis, Brazil
- NovemberRelease: continuous monitoring in grade A
- DecemberRelease: models for the cleanroom
The year in review is in the June 2025 news.