Calibration, maintenance and support for the whole TRIO.BAS range

Your TRIO.BAS always calibrated and ready to use: calibration, maintenance and support from a single partner.

Request a calibration → Our protocol

The ORUM laboratory in Milan
12 months
The calibration interval we recommend for TRIO.BAS, 6 months with heavy use
8 steps
For every instrument that comes into the laboratory
At your site
When needed, we calibrate at your premises
The ORUM laboratory in Milan

Calibration

Your instrument, calibrated by the people who designed it

We calibrate your TRIO.BAS samplers in the ORUM metrology laboratory in Milan.

Regulatory authorities ask for samplers used in pharmaceutical cleanrooms to be calibrated at regular intervals. We do it with a patented high-precision calibration system.

A new check is also needed when the instrument may have been damaged or its flow rate compromised.

Laboratory protocol

What we do on your instrument

Every TRIO.BAS that comes into the laboratory, for service or for its annual calibration, goes through these steps.

  1. 01

    Check for damage to the body

  2. 02

    Ultrasonic cleaning of the aspiration head

  3. 03

    Cleaning and check of the aspiration chamber

  4. 04

    Battery performance check

  5. 05

    Calibration and adjustment of the air flow rate

  6. 06

    Firmware update

  7. 07

    Replacement of damaged parts with original components

  8. 08

    New calibration certificate

The ORUM laboratory in Milan

Maintenance

The whole TRIO.BAS range, always in good working order

We look after the whole TRIO.BAS range, in our Milan laboratory or at authorised service centres.

Our protocol

TRIO.BAS

Commissioning

Ready to use from day one

We help you put the instrument into service and get the most out of it.

Write to us

FAQ

Calibration and maintenance, answered

The questions we are asked most often about TRIO.BAS service.

  • How often is an active sampler calibrated?

    Every 12 months, every 6 with heavy use.

  • What is the difference between calibration and adjustment?

    Calibration checks the deviation and produces a calibration certificate. Adjustment brings the instrument back into specification, typically within ±2.5%, and produces an adjustment certificate.

  • If the display says calibration expired, can sampling continue?

    The cycle can start. Results after the expiry may not stand in the official record. Book the calibration before it expires.

  • Does a new sampler need a new validation?

    Yes. A protocol exists for validating the new instrument.

  • Should the air flow be checked between two calibrations?

    Yes. If the flow is out of limits at the next calibration, the results of the whole previous period are not accepted. A check at regular intervals with a suitable instrument avoids this.

All questions on calibration and maintenance →

Support

In three steps

How to request service

  1. 1

    Write to us using the contact form and choose Calibration or Technical assistance.

  2. 2

    Give the instrument type and serial number, the problem and the action you need.

  3. 3

    We take it from there, in your country too.

Request a calibration
  • Technical support

    Calibration and repair with short turnaround, in Italy and abroad.

    Write to us

Does your TRIO.BAS need calibrating?

Write to us from the contact form with the instrument type and serial number: we will tell you timing and arrangements.

Request a calibration → Where we are →

Associations we are members of

  • S.I.M.A. (Società Italiana di Microbiologia Applicata)
  • ASCCA (Associazione per lo Studio e il Controllo della Contaminazione Ambientale)
  • PDA (Parenteral Drug Association)
  • A3P Association
  • ISPE (International Society for Pharmaceutical Engineering)